Anesthesia Machine
Anesthesia Machine Ventilator Anesthesia Spare Parts Anesthesia accessories Surgical lamp
7100B Plus Veterinary Anesthesia machine Specifications
Click here for online message:
Drop us a line
sales@jinxinmedical.com
Detail

Gas Supply:

Gas Source: O2,or O2&Air(Option), or O2&N2O(Option);                          

Gas Source Pressure Gauge: O2,or O2&Air(Option), or O2&N2O(Option); Pressure of Pipeline Inlet:                                                 

O2,280kPa600kPa/41Psi87Psi, less than 180L/min;N2O(Option),280kPa600kPa/41Psi87Psi,less than180L/min; Air(Option),280kPa600kPa/41Psi87Psi, less than 180L/min;Security Valve: release pressure less than 10 kpa;O2 Lost Alarm(Option):Sound alarm, alimumun cylinder with whistle;N2O cut off valve protection(Option): when O2 is lost, N2O will be cut off.


Flow Meter:

2 tubes high precision flow meter,01L/min,1.1L/min10L/min; or 0L/min10L/min(O2&Air or O2&N2O);Precision: At 20 degree, atmospheric pressure is 101.3kPa, the precision of flow meter is level:2.5;O2&N2O Linkage Device(O2&N2O,Option):make insure O2 concentration is not lower than 25%;


O2 Flush: 35L/min75L/min;


One Key Switch: Switch between rebreath and Non-Rebreath;


ACGO Port: Non-Rebreath;


Selectatec Mounting Bar: single place used for vaporizer, it is easy to replace;


Vaporizer:Anesthesia Vaporizer: one of Isoflurane/Halothane/Sevoflurane;


Respiration System/CO2 Circle Absorber/Rebreath:

Soda Lime Chamber: 800ml or 1600ml(Option);                                              

Pop Off Valve: with one key to close; APL valve(Option): 0.196kPa;Airway Pressure Gauge:-20-100cmH2O;


Anesthesia Ventilator

Drive gas: O2 or Air, Inlet pressure: 280-600 kPa. Max flow120 L / min;Safety Pressure: System pressure does not exceed 10 KPa;Display:10.1" TFT, with touch screen;Breath Mode:VCV,PCV,SIMV-VC,SIMV-PC,PSV,Manual/SPONT,SIMV-PRVC(option);Waveform:Pressure,Flow rate,Volume, Loop,PV loop, PF loop,FV loopO2 concentration Monitor:15%-100%(option);

Electric PEEP: OFF, 3 ~ 30 cmH2O;

Battery Type:Li-ion,work time more than 3 hours;


Parameters setting range:                        

Tidal volume:50 - 1500ml;                              

option :10 - 1500ml;                                     

Respiratory rate:1 ~ 100 bpm;                        

Inspiratorytime:0.1~10.0s;                                                  

Respiratory rate:0 ~ 100 bpm;Percentage of inspiratory pause:0 to 60%;    PEEP:OFF, 3 ~ 30 cmH2O;                           

Pressure Support:0 ~ 70 cmH2O; 

Pressure Control:5 ~ 70 cmH2O;                   

Flow trigger:0.5 ~ 20L/min;                            

Pressure Trigger:0 ~ 20 cmH2O;                  

PSV Insp Termination Level:5 ~ 80%; 


Monitored parameters:

Inspiratory tidal volume: 0 ~ 3000 mL;Expiratory tidal volume:0 ~ 3000 mL;Spontaneous minute ventilation:0 ~ 60 L / min;Minute ventilation:0 ~ 60 L / min;Respiratory ratio:4:1 to 1:10;

Spontaneous breathing frequency:0 ~ 100 bpm;

Compliance:0 ~ 300 mL/cmH2O;Respiratory ratio:30:1 to 1:150;Peak airway pressure:0 ~ 100 cmH2O;

Mean airway pressure:0 ~ 100 cmH2O;PEEP:0 ~ 100 cmH2O;Inspiratory plateau pressure:0 ~ 100 cmH2O;

FiO2(option):15 to 100%;EtCo2,FiCO2(option);

EtCO2, FiCO2, EtN2O, FiN2O, FiO2, EtO2, EtAA, FiAA,MAC(option);


Alarm settings:                                        

Tidal volume high/low alarm; Minute ventilation high/low alarm; Respiratory rate high/low alarm; Airway pressure high/low alarm; Continuous Positive Airway Pressure/Negative pressure alarm; Apnea alarm; Oxygen deficiency alarm; AC power failure alarm ,Low battery alarm.

 

Standard configuration:

Anesthesia machine, one unit X60 Anesthesia vaporizer, one set adult disposable breath circuit, one set pediatric disposable breath circuit, 0.5,1.2,3L airbag, O2, or O2&N2O, or O2&Air gas source pipeline

Option: SIMV-PRVC, O2 Concentration Monitor,EtCo2,FiCO2, EtCO2, FiCO2, EtN2O, FiN2O, FiO2, EtO2, EtAA, FiAA,MAC;


Dimension&Weight:Trolley Stand: 55cm*56cm*153cm,50kg;